Concept of the Week
New concept every Monday. Each week, one concept or development is explained in depth and in plain language for infection clinicians and laboratory microbiologists. Official documentation is linked when the source has been checked. Earlier concepts are listed in the Concept of the Week Archive.
Computer-aided detection (CAD) of tuberculosis on chest X-ray
In plain language. CAD for TB is software that looks at a digital chest X-ray and gives a score for how likely it is that the picture shows pulmonary tuberculosis. A person with a score above a chosen threshold is referred for further TB tests; a person below it usually is not. WHO has allowed this software to replace a human reader for TB screening and triage in people aged 15 and over since 2021, and in June 2025 it named six products that met its performance standards.a It is a triage tool, not a diagnosis. It works less well in some groups (for example people living with HIV and people treated for TB before), and the threshold has to be set for the local population.
Why this, why now
Two items in this week’s Journal Watch made me pick it. A medRxiv systematic review posted on 13 November 2026 pooled the accuracy of AI chest X-ray reading for TB in clinical cohorts,c and the SPOT-TB trial in The Lancet Global Health (November 2026) tested AI for choosing where mobile X-ray screening camps should go in Pakistan.d CAD for TB is also one of the few AI tools in infection practice with a named WHO recommendation behind it. So it is a good case study in what “evidence” for an AI tool should look like.
What it is and how it works
The software takes a plain, digital (DICOM) chest X-ray and returns a continuous abnormality score. The operator chooses a threshold score. Above it, the person is referred for further TB diagnostic evaluation. Below it, they are not.a (WHO policy statement 2025, p3.)
Here is the part that trips people up. Because the score is continuous, sensitivity and specificity are not fixed properties of the product. Move the threshold down and you catch more TB but refer more people who do not have it. Move it up and the reverse happens. WHO notes that some programmes favour sensitivity to maximise case detection while others favour specificity so as not to overwhelm laboratory capacity, and states that calibration is essential and that the right threshold depends on the programme’s goals, the hardware and the population.a (p3 and p12.)
Remember the scope. The WHO recommendation covers interpretation of plain digital chest X-rays for pulmonary TB in people aged 15 and over, for screening and triage. It does not cover children under 15, extrapulmonary TB, other image types, a differential diagnosis of chest disease, or confirming a TB diagnosis.a (p4.)
Where it is used in infection practice and the laboratory
- Community-wide and facility-based TB screening. These are the two use cases WHO assessed: screening people regardless of symptoms in the community, and screening or triaging people who attend a health facility.a (p1 and p4.)
- Triage before molecular testing. In a Dhaka cohort of 23,954 people, all five AI algorithms tested reduced the number of Xpert tests needed by half while keeping sensitivity above 90%.b Opinion For the laboratory, the practical point is that the threshold chosen shapes how many samples arrive.
- Settings short of trained readers. WHO describes the shortage of people able to read chest X-rays, and the variation between readers, as the main bottleneck that CAD is meant to ease.a (p2 to p3.)
- A related but different AI use: where to screen. SPOT-TB used AI software (MATCH-AI) to choose locations for active case-finding camps, not to read the X-rays themselves.d Keep the two ideas separate.
Routine use of TB CAD in UK NHS practice: not available from the sources I opened this week.
Evidence so far
| Source | Type | What it found |
|---|---|---|
| WHO, 2021 recommendation (as restated in the 2025 policy statement, p4)a | Guideline | Conditional recommendation that CAD may be used in place of human readers for TB screening and triage in people aged 15 and over. Certainty of evidence judged low. Based on three products evaluated in 2020. |
| WHO 2020 evaluation summary (2025 policy statement, Table 1, p3)a | Position statement | In community screening, CAD sensitivity was 90% to 92% with specificity 23% to 66%; human readers reached 82% to 93% sensitivity with 14% to 63% specificity. The ranges overlapped, so CAD and human readers were judged broadly similar. |
| WHO policy statement, June 2025 (p1 to p2, p6, Annex 3 p28 to p29)a | Position statement | 13 manufacturers applied; 5 products were excluded for not meeting recognised regulatory requirements; 8 were tested on independent image libraries (1,663 community, 2,252 facility images) against a microbiological reference standard. Six met both WHO criteria (partial AUC of at least 0.9 in the 40% to 60% specificity range, and at least 40% specificity at 90% sensitivity) in both settings; two did not. At 90% sensitivity, specificity ranged from 26% to 56% (community) and 30% to 49% (facility). Expert radiologists in the same libraries reached 94% sensitivity with 44% specificity (community). |
| Qin et al., Lancet Digital Health 2021b | Peer-reviewed | Five commercial algorithms on 23,954 chest X-rays from Bangladesh, all outperforming radiologists; only two met the WHO triage target (90% sensitivity, 70% specificity). Performance was worse in people over 60 and in people with previous TB. |
| Al Mahfuz et al., medRxiv, posted 13 November 2026c | Preprint | 21 studies; 9 evaluations from 3 cohorts pooled. Pooled sensitivity 0.896 and specificity 0.689. 18 of 24 assessed evaluations at high risk of bias; certainty low for sensitivity and very low for specificity. The authors state the search is incomplete. Not peer-reviewed. |
| SPOT-TB, Lancet Global Health, November 2026d | Peer-reviewed | Stepped wedge cluster-randomised trial, 3,936 camps, 269,254 people screened. AI-guided site selection: intention-to-treat risk ratio 1.07 (95% CI 0.94 to 1.22, not significant); per-protocol 1.32 (1.12 to 1.54). Tests targeting, not image reading. |
| Vendor or news claims | Vendor claim | None used. Vendor accuracy figures were not opened or relied on for this piece. |
Limitations, failure modes and bias
- Specificity is the weak point. At the 90% sensitivity WHO wants, specificity across the 2025 products fell well short of the 70% target product profile. WHO deliberately set its bar at 40%, close to expert radiologists.a (p11.) WHO itself notes the link between threshold choice and laboratory capacity (p12).
- Subgroup bias. Performance was lower in women, in people living with HIV and in people with previous TB, and lower in the oldest age group, though with wide confidence intervals.a (Annex 3, p29, p54, p69 to p70.) The FIND report suggests that under-representation of women in training data could explain part of the sex difference (p69). In the facility dataset, 86% of people had no HIV test result, so the HIV analysis is fragile (p70).
- Odd inputs. In error-handling tests, two of the eight products did not process about 45% of TB images containing a foreign body, most products scored non-chest images instead of flagging them, and six processed partial chest images without an error message.a (Annex 3, p29.) Where no human reader is present, that matters.
- Version drift. Only the exact versions reviewed by WHO count as meeting its standards; new versions may need re-evaluation.a (p12.) A product name alone tells you little.
- Thin real-world evidence. The 2026 preprint found that only 1 of 21 studies was prospective, fully verified, externally validated and met the WHO thresholds.c
- Non-TB findings. CAD for TB was not assessed for other abnormalities, yet people with such findings still need onward care.a (p12.)
Regulation and governance
- United Kingdom (MHRA). In the UK, many software and AI products are regulated as medical devices or IVDs. The MHRA’s guidance page (updated 3 February 2025) sets out its Software and AI as a Medical Device Change Programme (roadmap published 17 October 2022), the joint MHRA, FDA and Health Canada good machine learning practice principles, and Yellow Card reporting for software problems in England and Wales.e
- United Kingdom (NICE). The NICE Evidence standards framework for digital health technologies (ECD7, last reviewed 9 August 2022) was updated to include AI and adaptive algorithms. A NICE assessment specific to TB CAD: not available.
- WHO. WHO’s assessment covers diagnostic performance only, not product quality or manufacturing; countries must still meet local regulatory requirements. WHO intends its Prequalification Department to take on future, broader reviews.a (p2 and p12.)
- European Economic Area. The products assessed by FIND for WHO in 2025 were CE marked.a (Annex 3, p28.)
- Data. WHO calls integration with health information systems essential, including for data security of people being screened.a (p12.)
Practical takeaways for clinicians and laboratory staff
- Treat a CAD score as triage. It decides who goes on to further TB testing; it does not confirm a TB diagnosis. WHO excludes diagnosis confirmation from scope.a (p4.)
- Ask which product and which version, and whether that exact version is on the WHO 2025 list.a (p2 and p12.)
- Ask how the threshold was chosen and on which population. WHO says calibration is essential.a (p12.)
- Be more cautious in people living with HIV, people with previous TB, older adults and children (children under 15 are outside the WHO recommendation).a (p4, p13; Annex 3 p29.)
- Opinion If you run the laboratory, agree the threshold with the screening team in advance, because it sets your sample workload, and keep a log of CAD scores against culture or molecular results so local performance can be audited.
- Opinion In the UK, report software faults that affect patient care through the Yellow Card scheme (or the national equivalent), as for any medical device.e
Revising TB diagnostics for exams? Use the material on microregistrar.com rather than this page.
Glossary
- CAD
- Computer-aided detection: software that scores an image for a target abnormality.
- Triage test
- A test that decides who goes on to a confirmatory test.
- Threshold (cut-off)
- The score above which a person is referred. Changing it trades sensitivity against specificity.
- AUC and partial AUC
- Area under the ROC curve: overall discrimination. A partial AUC looks only at a chosen specificity range (WHO used 40% to 60%).
- Microbiological reference standard (MRS)
- In the WHO evaluation, a positive culture or Xpert/PCR on sputum within 14 days of the X-ray.
- Target product profile (TPP)
- WHO’s minimum performance wish list; for TB triage, at least 90% sensitivity and 70% specificity.
- SaMD / AIaMD
- Software as a medical device / AI as a medical device (MHRA terms).
References
- World Health Organization. Use of computer-aided detection software for tuberculosis screening: WHO policy statement. Geneva: WHO; June 2025. ISBN 978-92-4-011037-3. Section 1 (pp1 to 13) and Annex 3 FIND report (pp28 to 29, 54, 69 to 70). Guideline / position statement
- Qin ZZ, Ahmed S, Sarker MS, et al. Tuberculosis detection from chest x-rays for triaging in a high tuberculosis-burden setting: an evaluation of five artificial intelligence algorithms. Lancet Digit Health 2021;3(9):e543 to e554. Abstract, Findings. Peer-reviewed
- Al Mahfuz S, Roy SK, Rahman HI, Nahar A. Artificial intelligence for pulmonary tuberculosis detection on chest X-ray images: a systematic review and diagnostic accuracy meta-analysis. medRxiv, posted 13 November 2026. Abstract. Preprint, not peer-reviewed
- Mahfooz et al. Geographical targeting of active case-finding for tuberculosis in Pakistan using artificial intelligence software (SPOT-TB): a pragmatic, stepped wedge, cluster-randomised control trial. Lancet Glob Health, November 2026. Abstract, Findings (read via Europe PMC, PMID 43020100; page numbers not available). Peer-reviewed
- Medicines and Healthcare products Regulatory Agency. Software and artificial intelligence (AI) as a medical device. GOV.UK, updated 3 February 2025. Sections “Software and AI as a medical device change programme roadmap”, “Post-market and vigilance” and “Artificial intelligence”. Regulatory guidance
Model Context Protocol (MCP)
Model Context Protocol (MCP) is an open standard for connecting AI applications to external tools and data sources. Instead of writing a one-off integration for every system, a host application talks to MCP servers that expose tools, resources and prompts in a shared format.
For someone working in infection or the laboratory, the useful idea is practical rather than brand-specific: an assistant can be given controlled access to a database, a guideline library or a local tool, with the connection described in a standard way. This explainer is not a product endorsement, and it is not clinical advice.
Read the official MCP documentationEvery past concept is in the Concept of the Week Archive.
This site is for education and professional keeping-up. It is not clinical advice. Check the primary source before acting on any item. No patient-identifiable information is published here.